Please use this identifier to cite or link to this item: http://repository.kln.ac.lk/handle/123456789/4143
Title: Optical Design and Performance of the SASP Spectormeter at TRIUMF
Authors: Walden, P.L.
Walton, T.G.
Miller, C.A.
Yen, S.
Abegg, R.
Auld, E.G.
Campbell, J.R.
Chakhalyan, J.
Churchman, R.
Duncan, F.
Falk, W.R.
Frekers, D.
Green, A.
Green, P.E.W.
Hartig, M.
Hausser, O.
Haddock, C.
Hicks, K.
Hutccheon, D.
Jones, G.
Korkmaz, Y.Ke.E.J.
Khan, N.
Ling, A.
Lobb, D.E.
Naqvi, S.I.H.
Olsen, W.C.
Opper, A.
Otter, A.
Punyasena, M.A.
Reeve, P.
Zhao, J.
Keywords: Optics; Magnetic spectrometer
Sri Lanka; Magnetostatics; dispersion; Dispersion relations
Issue Date: 1999
Publisher: Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms
Abstract: A new Q?Q-Clamshell magnetic spectrometer called SASP has been commissioned in the proton hall at TRIUMF. The principal optical characteristics are: a solid angle acceptance of 13.5 msr, a momentum bite of 25%, a momentum resolution of ?p/p=2?10?4, and a maximum central momentum of 720 MeV/c. We describe the optical design, the magnetostatic design, the magnetic field mapping procedure, and the performance of this instrument.
URI: http://repository.kln.ac.lk/handle/123456789/4143
ISSN: 0168-583X
Appears in Collections:Physics

Files in This Item:
There are no files associated with this item.


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.